Changelog
V1.2.23
- Improved External Reference (ER) indicator:
- ER(red) – bad external reference signal or no external reference signal when it is expected (Timebase Reference set to External)
- ER! – external reference is detected but is not used (e.g. when Timebase Reference is set to Internal or if Timebase Reference is set to Auto and external reference is detected during ongoing measurement)
- ER – external reference is in use
V1.2.22
- Fixes for function/inputs select screen
- Improve settings switching time
V1.2.21
- Changes for on-screen keyboard
- CNT-102 support
- Added progress indication for SW update via USB stick
- Improvements for autoset
- Smarter control of rear fan
- Better overflow handling for Totalize
- Show Va/Vb auxiliary parameter for Phase measurement with auto-trigger
V1.2.17-p1
- Bug fix: Timebase Calibration was not applied properly on startup
V1.2.17
V1.2.12
- Fixes for Totalize and TIE functions
- Fixes for Arming
- Fix for voltage calibration in UI: first calibration point was rejected with an error
- Multiple fixes for SCPI
- Fixed unexpected buffer overflow errors
- Enhancements for Settings UI
V1.2.2
- Fix for timebase reference switching during measurement
V1.2.1