CNT-102 / 104S / 104R Firmware
Changelog
V1.2.27 (recommended)
- Improvements for Timebase Calibration UI
- Support for synchronizing 2 CNT-10x units using common arming (for e.g. timestamping 8 channels in parallel)
- Improved individual measurement speed via SCPI
- Improved AUTOSET
- File Manager: new hard to miss “Safe Remove” button for external USB storage
- SCPI: ’Function’ parameter of SYST:CONF command now allows spaces in inputs list
- Miscellaneous other improvements and fixes
V1.2.23
- Improved External Reference (ER) indicator:
- ER(red) – bad external reference signal or no external reference signal when it is expected (Timebase Reference set to External)
- ER! – external reference is detected but is not used (e.g. when Timebase Reference is set to Internal or if Timebase Reference is set to Auto and external reference is detected during ongoing measurement)
- ER – external reference is in use
V1.2.22
- Fixes for function/inputs select screen
- Improve settings switching time
V1.2.21
- Changes for on-screen keyboard
- CNT-102 support
- Added progress indication for firmware update update via USB stick
- Improvements for autoset
- Smarter control of rear fan
- Better overflow handling for Totalize
- Show Va/Vb auxiliary parameter for Phase measurement with auto-trigger
V1.2.17-p1
- Bug fix: Timebase Calibration was not applied properly on startup
V1.2.17
V1.2.12
- Fixes for Totalize and TIE functions
- Fixes for Arming
- Fix for voltage calibration in UI: first calibration point was rejected with an error
- Multiple fixes for SCPI
- Fixed unexpected buffer overflow errors
- Enhancements for Settings UI
V1.2.2
- Fix for timebase reference switching during measurement
V1.2.1
Download the firmware
V1.2.27 (recommended)
Old firmware versions
V1.2.23